[Home]
[Full version]
iPods and similar devices found not to affect pacemaker function
Mar 28 ,Medicine & Health
Last May, a widely reported study concluded that errant electronic noise from iPods can cause implantable cardiac pacemakers to malfunction. This just didn’t sound right to the cardiac electrophysiologists at Children’s Hospital Boston, who’ve seen hundreds of children, teens and young adults with heart conditions requiring pacemakers.
“Many of our pacemaker patients have iPods and other digital music players, and we’ve never seen any problem,” says Charles Berul, MD, director of the Pacemaker Service at Children’s. “But kids and parents bring up this concern all the time, prompting us to do our own study.”
Between September and December, 2007, Gregory Webster, MD, a cardiac fellow in training at Children’s, along with the electrophysiology nurses and physicians, ran tests on 51 patients coming in for appointments. Whereas last year’s study was done in patients averaging 77 years of age, the average age in the Children’s study was 22 (ranging from 6 to 60). All patients had active pacemakers or implantable cardioverter-defibrillators (ICDs), which were tested against four digital music players – two kinds of iPods (Apple Nano and Apple Video), SanDisk Sansa and Microsoft Zune. All patients were lying down during the tests (in case an arrhythmia occurred, causing them to faint), and each digital player was placed directly over the pacemaker or ICD.
As reported in the April 2008 issue of the journal Heart Rhythm, accompanied by an editorial, there was no interference with intrinsic device functioning – patients’ EKG (electrocardiographic) recordings showed no change in any of 255 separate tests, and no patients had symptoms. “This provides reassuring evidence that should allay the fears of people using iPods and other digital music players,” says Berul, the study’s senior investigator.
However, in 41 percent of patients, the music players interfered with telemetry, or communications between the programmer and the pacemaker or ICD itself. (The programmer is a computerized device used by physicians to check and recalibrate the pacemaker/ICD – patients do not carry it.) This interference, picked up in 29 of 204 tests, went away when the digital player was moved six inches or more from the device, and did not compromise device function. The larger digital players (Zune and Apple Video) caused more interference with telemetry than the small players.
Patients should not use digital music players while the doctor is trying to reprogram their device, Webster and the electrophysiology team conclude. “If the iPod is right in the field, the programmer might not be able to communicate with the pacemaker or ICD effectively,” Webster says.
The findings of interference with the pacemaker itself, reported last May, are published and updated in the same issue of Heart Rhythm. The editorial discusses possible reasons for the discrepancy between the two studies, which mainly relate to the testing methods and interpretation of the pacemaker recordings. Importantly, the music players did not stop pacemakers from pacing (regulating the heartbeat) in either the Children’s study or the previously reported study, though the previous study did find interference with pacemaker sensing.
Berul and colleagues are reassured by their own findings, but acknowledge that their testing was only short-term. “We can’t conclude that it’s completely safe to have an iPod right on top of the device for hours at a time,” Berul says. “That’s why we suggest the precaution of keeping it at least six inches away.”
Source: Children's Hospital Boston
Related stories:
Study: Hospital cell phone ban unnecessary
British researchers say they have determined the "hysteria" surrounding the use of cellular telephones in many hospitals is not justified.
Researchers find PDAs okay with pacemakers
With the dynamic evolution of wireless technology, Mayo Clinic researchers have been concerned about the potential effects of electromagnetic interference on heart pacemakers and implantable cardioverter-defibrillators. In the current issue of Mayo Clinic Proceedings, researchers report they did not detect interference from personal digital assistants (PDAs).
Device blocking stomach nerve signals shows promise in obesity
A new implantable medical device, developed in collaboration with Mayo Clinic researchers, shows promise as a reversible and less extreme alternative to existing bariatric surgeries, according to findings published in the current issue of the journal
Surgery.
Electromagnetic interference from some identification devices may pose hazards to medical equipment
The use of radio frequency identification devices appears to have the potential to cause critical care medical equipment to malfunction, according to a study in the June 25 issue of
JAMA.
Microchip sets low-power record with extreme sleep mode
A low-power microchip developed at the University of Michigan uses 30,000 times less power in sleep mode and 10 times less in active mode than comparable chips now on the market.
Video game technology may help surgeons operate on beating hearts
Surgery has been done inside some adults' hearts while the heart is still beating, avoiding the need to open the chest, stop the heart and put patients on cardiopulmonary bypass. But to perform intricate beating-heart operations in babies with congenital heart disease or do beating-heart complex repairs in adults, surgeons need fast, highly sophisticated real-time imaging that allows them to see depth. In an NIH-funded study featured on the cover of the June
Journal of Thoracic and Cardiovascular Surgery, cardiac surgeons from Children's Hospital Boston report good results with a simple technology borrowed from the gaming industry: stereo glasses.
Cardiac devices and advanced heart failure: Are we selecting the wrong patients?
Patients with advanced heart failure may be receiving implantable cardiac devices that do not help them because they are too ill to benefit from the treatment, a Saint Louis University study found.
Keeping the rhythm of life in sync
Beyond symbolically holding our feelings of love and compassion, the heart is a very efficient pump with a steady beat that provides the rhythm of life. Abnormal rhythm in the heart is a condition known as cardiac arrhythmia. A normal heart beats between 60 and 100 times per minute. It goes faster if needed for exercise or to handle emotional or physical stress.
[Home]